Dräger Babylog 8000 plus Instructions For Use Manual page 26

Intensive care ventilator for neonates software 5.n
Hide thumbs Also See for Babylog 8000 plus:
Table of Contents

Advertisement

Preparation
Calibration
Calibration:
Press the keys »Cal. Config.« and » -Cal«.
1
Remove the tube connector and seal the patient side of the
Y-piece with, for example, a sterile glove.
No gas may flow through the Y-piece during calibration
(since the zero point is calibrated).
Display:
2
Press the key »Start«.
After about 1 second, the display »
from the status window, indicating that the sensor has been
calibrated.
Display:
Connect the tube again.
If calibration is unsuccessful:
Replace the flow sensor insert or flow sensor cable,
page 27.
NOTE
If the flow sensor has to be replaced during operation, and
cannot be calibrated immediately, it should be noted that the
measuring accuracy may be reduced.
Recalibrate the flow sensor as soon as possible.
NOTE
If the cable of the flow sensor cable is temporarily
disconnected, it is not necessary to calibrate the sensor again.
26
Flow
« disappears
2
1

Advertisement

Table of Contents
loading

Table of Contents