CPAP with nasopharyngeal tube
Due to the leakage from the patient's mouth, it is not possible
to monitor the minute volume or to check for apnoea.
Therefore either switch off flow measurement:
Remove plug from Y-piece
or
set lower MV alarm limit to »0« and set apnoea time to
»OFF« (see under "Setting alarm limits" on page 50).
No monitoring of the lower minute volume and no apnoea
alarm activation.
Display:
Press the key »OK«. Flow measurement is now
switched off.
CAUTION
Note that the minute volume is no longer monitored and
there is no apnoea alarm!
Operation
CPAP