Fault – Cause – Remedy
Message
P
greater than 40 mbar?
insp
Press OK to confirm!
P
/PEEP
insp
Check set values!
Pressure measurement disturbed
Printer disturbed
printing cancelled
Re-calibrate flow sensor if
exchanged!
Tube obstructed?
VT low
Check settings!
80
Cause
The rotary knob for P
has been set to
insp
a value higher than 40 mbar, but the
setting is limited to 40 mbar.
P
has been set to less than 5 mbar
insp
above PEEP. PEEP is limited by P
Fluids in patient system.
Condensate in ventilation hoses.
Pressure measurement malfunction.
Internal diameter of ventilation hoses
too small.
Printer switched off.
No paper in printer.
Printer cable faulty.
RS 232 interface or printer configured
incorrectly.
During operation: flow sensor has been
replaced.
Cable has been disconnected and
re-connected.
Tube kinked or blocked.
Tidal volume does not reach the set value. Increase the flow
Remedy
Press »OK« to deactivate limiting to
40 mbar.
Increase P
, decrease PEEP.
insp
.
insp
Replace the patient system.
Remove the condensate.
Call DrägerService.
Use a suitable hose system.
Switch on the printer.
Insert paper.
Replace the cable.
Configure the RS 232 interface of the
printer and the Babylog 8000 plus with
identical settings.
Press »OK« and calibrate the sensor
(page 25).
Press »OK«.
Clear the tube.
Increase the inspiration time,
possibly P
.
insp
.
insp