Printing Reports - Dräger Babylog 8000 plus Instructions For Use Manual

Intensive care ventilator for neonates software 5.n
Hide thumbs Also See for Babylog 8000 plus:
Table of Contents

Advertisement

Press the key »Cal. Config.«:
Press the key »Print«.
Display (example):

Printing reports

This function is used for documenting measured values,
settings and the status of the ventilator.
Example:
To print a report once:
Press the key »Select« repeatedly until »Report« is
highlighted.
Start printing by pressing the key »Start«. The function of
this key then changes to »Stop«.
During printing of the report, the functions »All« and
»BabyLink« are not available. One of the other functions may
be started, but printing starts only after completion of the report
output.
To cancel the printing operation:
Press the key »Stop«.
Analogue and digital interfaces (optional)
+--------------------------------------------------------------------------------+
|15:39
23.08.1997
|
|Settings:
|
|TI:
0.40 s
Pinsp:
25.0 mbar
|TE:
0.80 s
PEEP:
4.1 mbar
|I:E: 1 : 2.0
FiO2:
21 %
|fset: 50 bpm
|
Flow in:
9.6 L/min
|
Flow ex:
9.6 L/min
+---------------------------------------+----------------------------------------+
|HFV:
|HfFreq: 10 Hz
HfAmpl:
100 %
|
|---------------------------------------+----------------------------------------+
|VG:
|VTset: 10.0 mL
|
+---------------------------------------+----------------------------------------+
|VIVE:
off
Trigger:
10.0
|
Operation
IPPV
Babylog 8000 plus
|
Measurements:
|
|
FiO2:
21 %
Mean:
10.1 mbar |
|
Leck:
0 %
Peak:
25.0 mbar |
|
spont:
0 %
PEEP:
6.0 mbar |
|
f:
50.0 bpm
|
VT:
15 mL
|
MV:
0,75 L/min
TIspont:
-.-- s
|
HFV:
|
VTim:
11 mL
MVim:
0.75 L/min |
|
VTHf:
--- mL
DCO2:
---- mL^2/s|
|
LFT:
|
C:
0.75 mL/mbar
R:
85 mbar/mL/s|
|
C20/C: 1.67
TC:
64 ms
|
Flow measurement:
|
NTPD
ISO
FiO2-compensated |
|
|
|
|
|
|
|
|
|
|
|
69