Testing Ippv; Testing Peep - Dräger Babylog 8000 plus Instructions For Use Manual

Intensive care ventilator for neonates software 5.n
Hide thumbs Also See for Babylog 8000 plus:
Table of Contents

Advertisement

Care
Checking readiness for use

Testing IPPV

Switch on IPPV/IMV (page 38).
Set the following rotary knobs (where the green LEDs are lit)
as follows:
1
»O
-Vol%« to 21.
2
2
»Insp. Flow
« to 10.
3
»T
« to 0.4
I
»T
« to 0.6
E
4
»P
« to 20
insp
5
»PEEP/CPAP« to 0
In the monitoring menu, press the keys »Meas« and »Paw«.
Display (example):
The displayed values must lie within their permissible
ranges, as follows:
Peak
18 to 22 mbar
Mean
6 to 10 mbar
PEEP
–1.5 to 1.5 mbar

Testing PEEP

6
Set the rotary knob »PEEP/CPAP« to 10.
Confirm this setting:
7
Press the key »OK«.
Display (example):
The displayed values must lie within their permissible
ranges, as follows:
Peak
18 to 22 mbar
Mean
12 to 16 mbar
PEEP
8 to 12 mbar
90
1
3
3
2
4
5
6
7