Volume guarantee VG (optional)

Volume guarantee VG (optional)
This can be combined with the ventilation modes SIPPV, SIMV
and PSV.
The inspiratory plateau pressure is automatically regulated
between P
and PEEP such that the set tidal volume VTset
insp
is applied.
VG requires a ventilation mode with a plateau.
This function is used for patients who are to be ventilated with
a constant tidal volume.
If the yellow LED blinks, consider:
The inspiration time is set too short. PSV/VG stroke has
ended.
The Tidal volume delivered is less than the VG tidal volume
set. The »VT low« alarm may activate.
Press the key »Vent. Option«.
Press the key »VG« and set the desired tidal volume VTset
with the keys »
« or »
Press the key »On«.
The tidal volume and the peak pressure of the regulated
ventilation stroke are displayed.
Adjust the rotary knobs »T
»Insp. Flow
« to obtain a ventilation pattern with
a plateau.
Adjust the rotary knob »P
value which is not to be exceeded.
When switching from a ventilation mode (SIPPV, SIMV or
PSV) with VG to IPPV, VG must be switched off.
Exit from the menu:
Press the key »
46
«.
«, »T
«, »PEEP/CPAP« and
I
E
« to the maximum permissible
insp
«.
Paw
P
insp
PEEP
TI
Flow
VTe
Vtrig
t
t
VTe
Vtrig