Note on EMC/ESD risk for the device function
General information on electromagnetic compatibility EMC/
ESD declaration according to international EMC standard
IEC 60601-1-2:
Medical electrical equipment requires special precautions
regarding electromagnetic compatibility (EMC) and needs to
be installed and put into service according to the EMC informa-
tion provided in the technical documentation. See Technical
Data, "EMC Declaration" on page 109.
Portable and mobile RF communications equipment can affect
medical electrical equipment.
WARNING
Pins of connectors identified with the ESD
warning symbol shall not be touched and
shall not be connected unless ESD precautio-
nary procedures are used. Such protective measures
may include antistatic clothing and shoes, touching
a ground stud before and during connection of the pins,
or using electrically insulating and antistatic gloves.
All personnel involved in the above shall receive instruc-
tion in these ESD protective measures .
Ensuring ventilation using an independent manual
ventilator
WARNING
In the case of an obvious defect in the device, the life-
support function may be endangered. The patient must
be ventilated immediately using an independent venti-
lator. If necessary, PEEP and/or an increased inspiratory
O
concentration should be applied (e. g. with the
2
manual ventilation bag MR-100).
For Your Safety and that of Your Patients
7