Testing Gas Failure Alarm - Dräger Babylog 8000 plus Instructions For Use Manual

Intensive care ventilator for neonates software 5.n
Hide thumbs Also See for Babylog 8000 plus:
Table of Contents

Advertisement

Testing gas failure alarm

Switch on the ventilator by pressing the main switch on the
rear of the unit until it locks into position.
Switch on IPPV/IMV (page 38).
Set lower alarm limit MV
MV
to 15 L/min.
1
Turn the rotary knob »O
Remove the oxygen connection hose from the wall outlet
socket.
2
The red alarm lamp must blink and the alarm must sound.
Display:
Re-insert the oxygen connection hose into the wall outlet
socket.
2
The red alarm lamp is extinguished, the alarm message
disappears and the alarm tone is switched off.
Remove the medical air connection hose from the wall outlet
socket.
Display:
2
The red alarm lamp must blink and the alarm must sound.
Re-insert the medical air connection hose into the wall outlet
socket.
2
The red alarm lamp is extinguished, the alarm message
disappears and the alarm tone is switched off.
to 0 and upper alarm limit
-Vol%« to 60 %.
2
Checking readiness for use
2
1
Care
89