Technical Data
EMC Declaration
Immunity against
Conducted RF
(IEC 61000-4-6)
*
For P
the highest possible "equivalent isotropic radiated power" of the adjacent RF transmitter has to be inserted
EIRP
(value in Watts). In the vicinity of equipment marked with the symbol
fixed, portable or mobile RF transmitters at the location of the medical device should be less than 3 V/m in the frequency
range from 150 kHz to 2.5 GHz and less than 1 V/m above 2.5 GHz.
**
ISM bands in this frequency range are: 6.765 MHz to 6.795 MHz; 13.553 MHz to 13.567 MHz; 26.957 MHz to 27.283 MHz;
40.66 MHz to 40.70 MHz.
Recommended separation distances to portable
and mobile RF telecommunication devices
The separation distances below are in accordance with
IEC 60601-1-2.
Max. P
(W) 150 kHz to 2.5 GHz
EIRP
0.03
0.32 m (1.05 ft)
0.10
0.58 m (1.90 ft)
0.17
0.76 m (2.49 ft)
0.20
0.82 m (2.69 ft)
0.25
0.92 m (3.02 ft)
0.41
1.18 m (3.87 ft)
0.82
1.67 m (5.48 ft)
1.00
1.84 m (6.04 ft)
1.64
2.36 m (7.74 ft)
3.28
3.33 m (10.93 ft)
Reduced separation distances to portable and
mobile RF telecommunication devices
The separation distances below are based on tests performed
by Dräger to determine the minimum separation distances.
These reduced separation distances are valid only for mobile
RF telecommunication devices using the standards listed.
T
Mobile communication device using ...
GSM 850, GSM 900, RFID 868 MHz (limited to 2 W ERP)
GSM 1800, GSM 1900 (limited to 1 W ERP)
UMTS, DECT (limited to 0.25 W ERP)
Bluetooth, WLAN 2450, RFID 2450 (limited to 0.1 W ERP)
112
Test level
(IEC 60601-1-2)
150 kHz to 80 MHz:
10 V inside ISM bands**
150 kHz to 80 MHz: 3 V
**)
outside ISM bands
All other frequencies Examples
0.96 m (3.15 ft)
1.75 m (5.74 ft)
2.28 m (7.48 ft)
2.47 m (8.10 ft)
2.76 m (9.06 ft)
3.53 m (11.58 ft)
5.00 m (16.40 ft)
5.52 m (18.11 ft)
7.07 m (23.20 ft)
10.00 m (32.81 ft)
Compliance level
Electromagnetic
(medical device)
environment
10 V
Recommended mini-
mum distance to porta-
ble and mobile RF trans-
3 V
mitters with transmission
power P
ical device including its
lines:
(1.84 m x √P
, interference may also occur. Field strengths from
e.g., WLAN 5250 / 5775 (Europe)
e.g., WLAN 2440 (Europe)
e.g., Bluetooth, RFID 2.5 GHz
e.g., WLAN 5250 (not in Europe)
e.g., UMTS mobiles
e.g., DECT devices
e.g., RFID 13.56 MHz
e.g., WLAN 5600 (not in Europe)
e.g., GSM 1800 / GSM 1900
e.g., GSM 900 mobiles, RFID 868 MHz
to the med-
EIRP
*)
)
EIRP
Separation distance
0.67 m (2.20 ft)
0.38 m (1.25 ft)
0.19 m (0.62 ft)
0.07 m (0.23 ft)