Advertisement

Replacing the insert of the flow sensor

This must be done if the following message is displayed:
Flow measurement disturbed
Measurement switched off
1
Disconnect the plug from the flow sensor.
2
Press the buttons on both sides and simultaneously pull the
insert out of the Y-piece.
3
Align the two marks and slide the new insert into the Y-piece
until it locks into position.
1
Connect the plug to the insert again, ensuring that the bar in
the insert engages with the slot in the plug.
Calibrate the flow sensor (page 25).
Preparation
Calibration
2
1
3
27