Operation
Displaying curves and measured values
Combinations of measured values
In the monitoring menu, press the keys »Meas« and
»MV O
P«.
2
Display (example):
MV
= expiratory minute volume
FiO
= measured inspiratory oxygen concentration
2
Mean = mean value of the airway pressure of the
preceding breathing cycle
Exit from the menu:
Press the key »
Displaying the measured volume values of high-frequency
ventilation (HFV)
In the monitoring menu, press the keys »Meas« and
»HFVol«.
Display (example):*
MVim = inspiratorially measured minute volume (L/min)
resulting from mandatory ventilation
DCO
= gas transport coefficient (mL
2
2
DCO
= VTHf
· f
2
f
= frequency [Hz] of the high-frequency pulses
VTim = inspiratorially measured tidal volume [mL] of
mandatory ventilation
VTHf = inspiratorially measured tidal volume [mL] of high-
frequency ventilation
Exit from the menu:
Press the key »
*
Only if the HFV option is installed.
See the Appendix, page 122, for a detailed description of HFV.
60
«.
2
/s)
«.