Accuracy - Mitsubishi Electric Q68AD-G User Manual

Channel isolated analog-digital converter module/ channel isolated analog-digital converter module (with signal conditioning function)
Table of Contents

Advertisement

3
SPECIFICATIONS
3.1.3

Accuracy

The reference accuracy is the accuracy at the ambient temperature for offset/gain setting.
The temperature coefficient is the accuracy per temperature variation of 1°C.
The reference accuracy is the accuracy relative to the maximum digital output value.
Even if you change the offset/gain setting or input range to change the input characteristic,
the reference accuracy and temperature coefficient do not change and are kept within the
ranges given in the performance specifications.
[Example] Accuracy when the temperature changed from 25°C to 30°C
0.1% (reference accuracy) + 0.00714%/°C (temperature coefficient) 5°C (temperature
variation difference) = 0.1357%
3.1 Performance Specifications
3.1.3 Accuracy
1
2
3
4
5
6
7
8
3
- 12