Accuracy Specifications Of The Current Probes - AEMC 8230 User Manual

Powerpad jr. single-phase power quality analyzer
Table of Contents

Advertisement

3.2.4

Accuracy Specifications of the Current Probes

These characteristics are stated after linearization. The errors of the sensors are
compensated by a typical correction inside the instrument. This typical correction
is in phase and in amplitude according to the type of sensor connected (detected
automatically) and the gain of the current acquisition chain used.
The measurement error in RMS current and the phase error are additional errors
(they must therefore be added to those of the instrument alone) stated as influ-
ences on the calculations performed by the analyzer (powers, energies, power
factors, tangent, etc.).
Type of sensor
MR193
1000A
SR193
1000A
AmpFlex
®
3000A
MiniFlex
®
1000A
MN93
200A
MN193
100A
MN193
5A
5A
Adapter
*DC Only
16
TRMS current
[1A; 10A]
[10A; 100A]
[100A; 800A]
[800A; 1200A]
[1200A; 1400A]*
[1A; 3A]
[3A; 10A]
[10A; 100A]
[100A; 1200A]
[10A; 100A]
[100A; 6500A]
[10A; 100A]
[100A; 1000A]
[0,5A; 2A]
[2A; 10A]
[10A; 100A]
[100A; 240A]
[100mA; 300mA]
[300mA; 1A]
[1A; 120A]
[5mA; 50mA]
[50mA; 500mA]
[500mA; 6A]
[5mA; 50mA]
[50mA; 6A]
Maximum error on IRMS
±(1.5%+1A)
±(3%)
±(5%)
±(0.8%)
±(0.3%)
±(0.2%)
±(3%)
±(2%)
±(3%)
±(2%)
±(3%+1A)
±(2.5%+1A)
±(1%+1A)
±(0.7%+2mA)
±(0.7%)
±(1%+0.1mA)
±(1%)
±(0.7%)
±(1%)
±(0.5%)
Power Quality Analyzer Model 8230
φ
Maximum error on
N.S.
±(2°)
±(1.5°)
N.S.
±(1°)
±(0.5°)
±(0.3°)
±(1°)
±(0.5°)
±(1°)
±(0.5°)
N.S.
±(6°)
±(3°)
±(2°)
N.S.
±(1.5°)
±(0.7°)
±(1.7°)
±(1°)
±(1°)
±(0°)

Advertisement