AEMC 8230 User Manual page 53

Powerpad jr. single-phase power quality analyzer
Table of Contents

Advertisement

3.
Define the measurement values, then press the Enter button.
Use the ▲▼ buttons to select a field, press Enter to enter the field, ▲▼ to incre-
ment or decrement the value in the field, ◄► to go from one selection to the other
inside the field, and press Enter to exit from the edit mode.
• Start Threshold: defines the threshold beyond which recording will be
affected. (Note: The triggering and stopping thresholds are half-period
RMS current values.)
• Hysteresis: determines, with the triggering threshold, the stopping
threshold of Inrush recording. With a hysteresis of 2% and a triggering
threshold of 1000A
(Note: When the stopping threshold is not detected, recording continues
until the memory is full. The maximum value is 5999A
• Start: defines the date and time in which the recording will be done.
4.
Select OK and press Enter to validate. The message "Waiting for triggering
threshold" will display at the bottom of the screen.
When the triggering threshold is reached, the current measurements Inrush
will be recorded. Recording will stop when the stopping threshold is reached.
NOTE: To stop the recording at any time, select the Hand tool (
the ▼ button and then the Enter button.
Power Quality Analyzer Model 8230
NEW INRUSH
: 0100Arms
Start Threshold
: 2%
Hysteresis
: 02/25/06 15:20
Start
, the stopping threshold will be 980A
RMS
02/25/06 10:26
OK
Figure 5-23
.
RMS
.)
RMS
) by pressing