Entering Transducer and
Pipe Parameters - CHx-
PIPE
Accessing the PIPE
Submenu
Programming Site Data
Enter the transducer and pipe parameters via the
following the programming instructions, refer to Figure A-1 on page
A-1 in Appendix A, Menu Maps. Remember to record all
programmed data in Appendix C, Data Records.
1. Press [ESC], [ENTER], [ESC]. The meter enters the user program.
2. Press [ ] until PROG appears and press [ENTER].
3. Press [ ] until the desired channel appears. Press [ENTER] at CH1
or CH2.
4. Press [ ] until PIPE appears and press [ENTER].
5. Do one of the following:
•
Standard transducers, enter the transducer number by using the
arrow keys. Press [
location and press
number and press [ENTER]. Proceed to the next step.
•
Special transducers, proceed to Special Transducers below.
IMPORTANT:
Special transducers have no engraved number on the
head and are rarely used. Examine the transducer
head carefully for a number.
6. Go to one of the following sections:
•
Standard clamp-on transducers - proceed to Pipe Material on
page 2-7.
•
Standard wetted transducers - proceed to Pipe Outside
Diameter on page 2-8.
PIPE
]
and [ ] to move the cursor to desired
[ ] and [ ]
to increase or decrease the
September 2004
submenu. While
2-5