4 GETTING STARTED (continued)
To fit / replace the probe tip:
1
Unscrew the probe tip collar (a)
from the probe body (b).
2
Remove the worn or damaged
probe tip.
3
Fit a new probe tip (c) by sliding it
into the probe tip collar.
4
Apply a small drop of probe tip oil to
the sensor plate (d).
5
Refit the probe tip collar to the
probe body.

4.6 OILING THE SENSOR PLATE

For the probe to work correctly, there must be no air gap between the
probe tip and sensor plate. This is achieved by using a small amount
of probe tip oil (supplied with each gauge ). The gauge will inform the
user when oil is required.
To oil the sensor plate:
1
Unscrew the probe tip collar (a)
from the probe body (b) and
remove the probe tip (c) from the
sensor plate (d).
2
Apply a small drop of probe tip oil to
the sensor plate (d).
3
Refit the probe tip and probe tip
collar to the probe body.
Note: The use of normal oil is not recommended as it could damage the probe tip and
affect the accuracy of the gauge. Ultrasonic couplant can be used as an
alternative however, the probe tip and sensor plate will need to be cleaned more
regularly and couplant re-applied more frequently.
Note: A Material Safety Data Sheet for the probe tip oil supplied by Elcometer is
available to download via our website:
www.elcometer.com/images/stories/MSDS/Elcometer_Probe_Tip_Oil.pdf
a
Additional probe tips and probe tip oil can be purchased from Elcometer or your local Elcometer supplier,
see Section 20.1 'Probes' on page en-31.
en-7