Set Time And Date - Smiths Medical CADD-Solis Operator's Manual

Ambulatory infusion pump
Hide thumbs Also See for CADD-Solis:
Table of Contents

Advertisement

WARNING: Ensure that the entire fluid path is
free of all air bubbles before connecting to the
patient to prevent air embolism. Air embolism
could result in serious patient injury or death.
NOTE: If the fluid path contains an air
eliminating filter, it is acceptable for air bubbles
to be present on the vent side of the filter. See
the cassette's instructions for use for more
information.

Set Time and Date

The set time and date screen allows you to edit the
time and date, as well as to choose a date format.
The date format options are Month/Day/Year or
Day/Month/Year.
The time option should reflect the current time.
The pump shows the time of day in 24-hour time
according to the pump's internal clock. The clock is
powered by a separate, internal battery which retains
the time even when the 4 AA batteries or the battery
pack is removed. The pump uses its clock to record
the time of events in the delivery and event logs, as
well as in other reports.
The date option should reflect the current date. The
pump uses this feature to record the date of events
in the delivery and event logs, in other reports, and
to determine when the preventative maintenance
(PM) reminder alarms will occur.
NOTE: Protocol information will remain the same
regardless of changing the time and/or date during a
patient's therapy.
The pump must be stopped and you'll need the
keypad code (or a higher level code) or the cassette/
keypad lock must be unlocked.
To access the set time and date feature from the
tasks menu:
1. Press U or D to
choose Set Time and
Date and press S.
Programming and Operation
To set the current time:
1. Press U or D
until Current Time is
highlighted and press
S.
2. Unlock the keypad using
the security code or the pump key.
3. Press U or D to
scroll to the correct
hour and press S to
navigate to the minutes.
4. Press U or D to
scroll to the correct
minutes and select Save.
Updating the time and date to daylight savings time.
The pump's time and date will not automatically
update during daylight savings. If you live in a
geographical area that follows daylight savings time,
you will need to manually update the time and date.
You may choose to change the time while the pump
is currently being used by a patient, or to wait until
the patient is finished with therapy.
NOTE: If you update the pump's time while it is
being used on a patient, the timestamps in the event
and delivery logs will not be updated to reflect
daylight savings prior to the change. All events
will record the reported time from when the event
actually occurred. For your reference, the event log
will record the time it was changed.
To set the current date:
1. Press U or D
until Current Date is
highlighted and press
S.
2. Unlock the keypad
using the security code or the pump key.
3. Press U or D to scroll to the correct month
and press S to navigate to the day.
Press here to Save
25