Specifications (Nominal); General Pump Specifications - Smiths Medical CADD-Solis Operator's Manual

Ambulatory infusion pump
Hide thumbs Also See for CADD-Solis:
Table of Contents

Advertisement

References and Troubleshooting
EN 55011 (1998), Industrial, scientific and
medical (ISM) radio frequency equipment
- Radio disturbance characteristics - Limits
and methods of measurement. Amendment 1
(1999), Amendment 2 (2002)
Miscellaneous Standards
USB 1.1 Universal Serial Bus (USB)
Specification, Revision 1.1, September 23,
1998 – USB.org
EN 1041 (1998), Information supplied by the
manufacturer with medical devices.
IEC/TR 60878 (2003), Graphical symbols for
electrical equipment in medical practice.
EN 980 (2003), Graphical symbols for use in
the labeling of medical devices.
IEC 60529 (2001), Degrees of protection
provided by enclosures (IP Code).

Specifications (Nominal)

General Pump Specifications

Medication cassette reservoir, part number 21-7002,
®
CADD
extension sets, part number 21-7045, and
®
CADD
administration sets, part number 21-7091
were used to test the pump.
Resolution
Medication cassette reservoir or CADD
administration set, 0.050 mL per pump stroke
nominal.
Size
4.1 cm × 10.2 cm × 12.7 cm [1.6 in. × 4 in. ×
5 in.] excluding cassette or other accessories.
Weight
595 g [21 oz.] including 4 AA alkaline batteries,
excluding other accessories.
Pump Alarms
The following alarms are all considered to be
high priority alarms: Battery depleted while
delivering; Battery removed while delivering;
Battery unusable while delivering; Rechargeable
battery end of life; Disposable detached while
delivering; Disposable attached improperly;
Disposable type high flow administration set;
46
Disposable type invalid; Disposable locked but
not latched; Reservoir volume empty; Pressure
sensor faulty; Downstream occlusion; Upstream
occlusion; Air in line detected; Remote dose
cord key stuck; Key stuck; Stop mode reminder;
Pump was automatically stopped.
There are an additional 19 medium
priority alarms, 8 low priority alarms, and
20 informational messages alerts.
Battery Fallout Alarm
Alarm sounds for 2 minutes if the pump has
been powered up for a minimum of 2 minutes.
NOTE: Only enabled while pump is in run
mode.
Battery Status
Battery State
25%—100%
Low battery
Depleted
battery
Shut down
CADD®-Solis pump low battery conditions.
®
The pump emits 3 beeps every 5 minutes, and the low battery warning
message appears on the pump's display, indicating that the battery
power is low, but the pump is operable.
††
The pump emits a continuous, variable-tone alarm, and the depleted
battery warning message appears on the display, the battery power is
too low to operate the pump, and the pump delivery operation has
stopped.
Continuous Delivery Alkaline Battery Life
with screen intensity set to 3
0.4 mL/hr
1 mL/hr
5 mL/hr
10 mL/hr
30 mL/hr
CADD® Pump Status
No alarm
Transition to low battery
condition; battery low message
appears; 3 beeps every 5 min.
LCD backlight will flash for 12 ms
during each motor operation.
Transition to depleted battery
condition; battery depleted
message appears; pump beeps
continuously.
††
Pump shuts off due to too low
operating voltage.
Rate
Life
260 hrs
210 hrs
165 hrs
150 hrs
70 hrs
Volume
104 mL
210 mL
825 mL
1500 mL
2100 mL