Smiths Medical CADD Solis VIP Step-By-Step Manual

Smiths Medical CADD Solis VIP Step-By-Step Manual

Manual programming for continuous therapy mode
Hide thumbs Also See for CADD Solis VIP:

Advertisement

Quick Links

NOTE: The following video demonstrates the steps required to program a CADD
Pump in Manual Mode. The steps used are the same as those used to program a CADD
Infusion Pump in Manual Mode. CADD
from the library saved on the pump to program each infusion.
CADD® Solis VIP
Solis VIP
Step by Step Clinician Guide:
Step by Step Clinician Guide:
Manual Programming
for Continuous Therapy Mode
for Continuous Therapy Mode
-Solis VIP with Select Library or PharmGuard Enabled use a drug protocol
®
-Solis VIP v1.3 Ambulatory Infusion
®
-Solis VIP v1.5 Ambulatory
®

Advertisement

loading
Need help?

Need help?

Do you have a question about the CADD Solis VIP and is the answer not in the manual?

Questions and answers

Summary of Contents for Smiths Medical CADD Solis VIP

  • Page 1 CADD® Solis VIP Solis VIP Step by Step Clinician Guide: Step by Step Clinician Guide: Manual Programming for Continuous Therapy Mode for Continuous Therapy Mode NOTE: The following video demonstrates the steps required to program a CADD -Solis VIP v1.3 Ambulatory Infusion ®...
  • Page 2 Pump displays the home screen. Press Tasks. Scroll down to View Advanced Tasks Press Select Scroll down and select Start New Patient...
  • Page 3 Press Continue Enter unlock code and press Accept Value Refer to your organization for the security code Scroll down and select Continuous Pump begins programming sequence...
  • Page 4 Press Review Press select to edit Continuous Rate parameter Once desired setting is confirmed, press Save Scroll up or down to adjust infusion rate...
  • Page 5 Press Accept Value to confirm parameter Press select to edit Reservoir Volume Adjust Reservoir volume Once desired setting is confirmed, press Save...
  • Page 6 After confirming all parameters, press Next Attach cassette or admin set to pump Press Yes to start prime sequence if applicable Press Prime to start priming...
  • Page 7 Press Stop Priming once air removed from the line Press No to end the prime sequence Press Yes to start the infusion Pump is running...
  • Page 8 To change rates during infusion, select stop Press Yes Pump stopping Press Program...
  • Page 9 Press select to edit Continuous Rate parameter Enter security code and press Accept Value Adjust infusion rate and press Save Press start to review pump settings...
  • Page 10 Press Review Press Accept Value to confirm parameters After confirming all parameters, press Next Press Yes to start the pump...
  • Page 12 Smiths Medical is part of the global technology business Smiths Group plc. Please see the Instructions for Use/Operator’s Manual for a complete listing of the indications, contraindications, warnings and precautions. CADD, CADD Solis and the Smiths Medical design mark are trademarks of Smiths Medical. The symbol ® indicates the trademark is registered in the U.S. Patent and Trademark Office and certain other countries.