a clinician bolus, it will give you the option to restart
the clinician bolus where it left off or to start with a
new clinician bolus.
WARNING: Exercise care when using the clinician
bolus function. Since there are no limits on the
frequency of delivering a bolus, and since the
amount of the bolus can be set as high as 20
mL (or the mg or mcg equivalent), you should
not permit the patient to become familiar with
the procedure for giving a clinician bolus.
Improper programming could result in serious
patient injury or death.
NOTE: The maximum clinician bolus may be limited
by the settings in the protocol, which is determined
®
by the CADD
-Solis system administrator.
To start a clinician bolus:
1. Make sure the pump is running. Start the pump
if necessary.
2. From the Tasks menu,
press U or D until
Give Clinician Bolus is
highlighted, then press
S.
3. Unlock the keypad using
the pump key (if allowed)
or press U or D to
enter the clinician code
(or a higher level code).
Press S to advance to
the next digit. Once the
code has been entered,
select Accept Value.
NOTE: If you enter a
code that you believe
is correct and receive
a wrong code error,
check the screen to see
if the clinician code
has been customized.
If the code has not been customized, it was
entered incorrectly. Retry. If the code has been
customized, use the custom clinician code.
If you do not know this code, contact your
®
CADD
-Solis system administrator.
4. Make sure the clinician
5. The screen will show the

Start New Patient

Press here to Accept Value
Each time a new patient is started, it will be
recorded in the event log. All other reports will be
cleared.
The pump must be stopped and you'll need the
keypad code (or a higher level code) or the cassette/
keypad lock must be unlocked, if allowed. To start a
new patient:
1. From the Tasks menu,
Programming and Operation
WARNING: To prevent the patient from
accessing the clinician bolus function, do
not let the patient know the security codes.
Improper programming could result in
serious patient injury or death.
bolus amount is at the
desired value and select
Deliver.
NOTE: If you enter a
value outside the soft
limit, a screen will
appear asking you to
confirm the soft limit
override.
amount decreasing as the
bolus is delivered. You
may stop the bolus at any
time by selecting Stop
Bolus.
WARNING: Never leave the pump unattended
while on the clinician bolus edit screen. You
must press "Confirm" or "Deliver" to deliver
the programmed value or cancel to leave the
screen. Failure to do so could result in serious
patient injury or death.
press U or D until
Start New Patient is
highlighted and press
S.
Press here to Deliver
Press here to Stop Bolus
23