Appendix B: Setting the Optional Real Time Clock

1.
Clear trend memory by pressing and holding the print key at power up until CLR stops
flashing.
2.
Turn the monitor off. Power on the monitor and continue to hold the on key for 6-8 seconds.
The version number should be displayed and then the display will go blank except for one set
of digits. This is the minutes setting.
3.
Adjust the minutes by pressing the print key until the desired setting is reached.
4.
Press the power key again to show the hour setting. Use the print key to set the desired hour.
5.
Press the power key again to show the day setting. Use the print key to set the desired day.
6.
Press the power key again to show the month setting. Use the print key to set the desired
month.
7.
Press the power key again to show the year setting. Use the print key to set the desired year.
8.
Press the power key again to exit the time setting menu. After the power key is pressed a
time stamp is printed and the monitor turns itself off.
9.
Verify that the correct time stamp was printed. If the correct time and date was not printed,
repeat steps 2 through 9 until the proper time and date is verified.
The time and date can be verified by repeating step 2 and then pressing the print key until the
current time and date is printed and the monitor turns itself off.
Pulse Oximeter for Sleep Screening Operation Manual
Appendix B
B-1