Start New Protocol, Same Patient; Prime Tubing - Smiths Medical CADD-Solis Operator's Manual

Ambulatory infusion pump
Hide thumbs Also See for CADD-Solis:
Table of Contents

Advertisement

Programming and Operation
2. Unlock the keypad using the security code or
the pump key.
3. Press U or D to
highlight the desired
therapy and press S.
4. Press U or D to
highlight the desired
qualifier and press S.
5. Press U or D to
highlight the desired
drug and concentration,
and press S.
6. The new protocol will
display on the screen and
the pump will ask you
if you have chosen the
correct Therapy, Qualifier,
Drug, and Concentration.
Confirm that you have
selected the correct
Therapy, Qualifier, Drug,
and Concentration or Units.
NOTE: If programming in mL, you will not be
asked to confirm the concentration.
Selecting Yes will program the pump with the
protocol you have chosen.
Selecting No will bring you back to the Select
Drug screen. If you desire to change the drug,
repeat the process for selecting a new drug.
Otherwise, select Back until you reach the screen
you desire to change (therapy or qualifier).

Start New Protocol, Same Patient

The process for starting a new protocol for the same
patient is much like starting a new patient; however,
the event log does not insert a new patient marker
(all other reports will be cleared, except the delivery
log). See Start New Patient on page 23 for directions
24
on selecting a therapy, qualifier, and drug for the
new protocol.
NOTE: When starting a new protocol, be sure to
attach a new reservoir with the proper drug and
concentration.

Prime Tubing

Priming the tubing is done to fill the tubing
downstream of the pump with fluid, removing any
air bubbles. Prime the tubing before connecting it to
the patient's infusion set or indwelling catheter. The
pump must be stopped and you'll need the keypad
code (or a higher level code) or the cassette/keypad
lock must be unlocked.
NOTE: The air detector is disabled while the pump
is priming.
WARNING: Do not prime the fluid path with
the tubing connected to a patient as this could
result in overdelivery of medication or air
embolism, which could result in serious patient
injury or death.
To prime the tubing:
1. From the Tasks menu,
press U or D until
Prime Tubing is
Press here to select
highlighted and press
S.
2. Unlock the keypad using the security code or
the pump key.
3. If you have not already
done so, disconnect the
tubing from the patient,
open clamps, and select
Prime.
You may stop priming
at any time by selecting
Stop Priming. Otherwise
priming will automatically
stop once it has primed
10 mL. Continue priming
as needed.
Press here to select
Press here to stop