Pump Diagram

Front View

Indicator Lights

Amber
Green
USB Port
Blue AC
Power Light
AC Power Jack
Remote
Dose Cord
Jack
Keypad
Rear View
Power Switch
Cassette/Keypad Lock
(Unlock/Lock)
Cassette Latch
Cassette (the part
of the medication
cassette reservoir
or CADD®
administration set
that attaches to
the pump)
Battery Compartment
Display
Description of the Keys,
Components, Ports, and Connectors
Indicator Lights
When the pump is powered, one or both of the
indicator lights will flash.
Green: The green light flashes to indicate
that the pump is running and delivering fluid
as programmed.
Amber: The amber light flashes when the
pump is stopped, an alarm condition exists,
or the battery or the reservoir volume is low.
It stays on continuously when the pump is
inoperable. The display will briefly describe
the alarm condition when the amber light is
flashing.
NOTE: At times both lights may flash. This
indicates that the pump is running, but there
is a condition the clinician should be aware of
(e.g., low battery or low reservoir volume).

Display with backlighting

The liquid crystal display (LCD) shows
programming information and messages.
Backlighting helps keep the display visible in low
light. In this manual, "display" is synonymous with
display panel or LCD.
After a period in which no keys are pressed, the
backlighting turns off and the display goes blank
to save battery power (except during an alarm or
when an external power source is in use). You may
press any key to turn the display back on.
NOTE: When the display is blank, you can
determine that the pump is powered by observing
either the green or amber (or both) LED indicators
periodically flashing.
NOTE: If you press stop/start 9, the display will
reappear with a message asking if you wish to start
or stop the pump.
NOTE: If you press PCA dose @ while the pump
screen is blank, the pump will deliver a PCA dose,
if available.