Delivery Specifications; Administrator Settings Specifications - Smiths Medical CADD-Solis Operator's Manual

Ambulatory infusion pump
Hide thumbs Also See for CADD-Solis:
Table of Contents

Advertisement

References and Troubleshooting
High Pressure Alarm Threshold
18 ± 9 psi [1.24 ± 0.62 bar].
Air Detector Alarm
Sensitivity:
Low - Single bubble greater than 400 μL .
High - Single bubble greater than 150 μL .
Accumulated Air: Greater than 1 mL air over
15 minutes (nominal).
Bolus Accuracy Specification ± 6%
Actual Test Data for Bolus Accuracy at 0.05 mL:
Average
% Error
Minimum Error %
Maximum Error %
Actual Test Data for Bolus Accuracy at 20 mL:
Average
% Error
Minimum Error %
Maximum Error %
Maximum Volume Infused Under Single-Fault
Conditions
®
CADD
administration set: 0.15 mL.
Delivery Rate during priming
Approx. 250 mL/hr.
Alarm disabled during priming
Air-In Line.

Delivery Specifications

Reservoir Volume
0 to 9999; programmable in 1 mL increments,
displayed in 0.1 mL increments.
Programming Units
Milliliters (mL), milligrams (mg), micrograms
(mcg).*
*If programming through the manual mode.
Otherwise the programming units are preset
through the CADD
Software.
48
0.0508 mL
1.6%
–3.0%
4.2%
20.108 mL
0.5%
–2.9%
4.5%
®
-Solis Medication Safety
Concentration
mg/mL:
0.1 to 0.5 mg/mL in increments of 0.1 mg/mL.
0.5 to 1 mg/mL in increments of 0.5 mg/mL.
1 to 15 mg/mL in increments of 1 mg/mL.
15 to 100 mg/mL in increments of 5 mg/mL.
mcg/mL:
1 to 15 mcg/mL in increments of 1 mcg/mL.
15 to 100 mcg/mL in increments of 5 mcg/mL.
00 to 500 mcg/mL in increments of 100 mcg/mL.
Continuous Rate
0 to 30 mL/hr (or the mg or mcg equivalent).
PCA Dose
0 mL to 20 mL (or the mg or mcg equivalent).
Delivery rate (continuous rate + PCA dose):
programmable from 40 to 175 mL/hr.
PCA Dose Lockout
1 minute to 24 hours in the following
increments:
1 minute for values between 1 and 20 minutes.
5 minutes between 20 minutes and 24 hours.
Max Doses per Hour
1 to 60.
Delivery Limit Amount
0.1 to 1000 mL (or the mg or mcg equivalent)
in increments of:
0.01 mL from 0.01 to 0.5 mL.
0.5 mL from 0.5 to 100 mL.
1.0 mL from 100 to 1,000 mL.
Given
0 to 99,999.99 in 0.01 unit increments.
Clinician Bolus
0 mL to 20 mL (or mg or mcg equivalent)
delivery rate (continuous rate + clinician bolus):
programmable from 40 to 175 mL/hr.

Administrator Settings Specifications

Delivery Limit Method
Delivery limit, max doses per hour, or not in
use.
Delivery Limit Period
1 to 12 hours in increments of 1 hour.