3. Press U or D until
you reach the desired
PCA lockout time and
select Save.
NOTES:
If the delivery limit is reached while a PCA
dose is in progress, the PCA dose will not be
completed.
A PCA dose cannot be started while another
PCA dose or a clinician bolus is in progress.
Pressing the remote dose cord button will turn
the display back on AND deliver a PCA dose
(if available). A blank display will not require
two key presses to start a PCA dose.

Hourly Limit

The hourly limit is the amount of drug which can be
delivered in a specified time frame (1-12 hours). The
time frame is determined by your facility's CADD
Solis system administrator. This limit includes
the continuous rate and PCA doses, but does not
include clinician boluses.
If the delivery limit is reached and a continuous rate
is programmed (value other than zero), the status
bar will display "Delivery Limit" and the pump will
deliver a KVO rate of 0.1 mL/hr. If the delivery limit
is reached and no continuous rate is programmed
(value = zero), the status bar will display "KVO = 0"
and the KVO rate will be zero.
NOTE: If the delivery limit is reached while a
PCA dose is in progress, the PCA dose will not be
completed.
The hourly limit can be edited while the pump is
running or stopped (if the keypad is unlocked).
To edit the hourly limit:
1. From the Home screen,
press U or D to
choose X* Hour Limit
and press S.
*X will be the number of hours set up by the CADD®-Solis system
administrator. In the example shown, a 4 hour limit is being used.
2. Unlock the keypad
using the security code
or the pump key.
3. Press U or D until
you reach the desired
hourly limit amount
and select Save.
Press here to Save
If the hourly limit is not visible on the home screen,
your CADD
not to require it. Instead you may be required to
program the Max Doses/Hr.

Max Doses/Hr

If the max doses/hr setting is turned on, you may
use it to further restrict the number of PCA doses
available to the patient in 1 hour. The max doses/hr
feature allows you to restrict the PCA doses beyond
the PCA lockout time.
The max doses/hr time can be edited while the
pump is running or stopped (if the keypad is
unlocked).
®
-
To edit the max doses/hr:
1. From the Home screen,
press U or D to
choose Max Doses/Hr
and press S.
2. Unlock the keypad
using the security code
or the pump key.
3. Press U or D until
you reach the desired
max doses/hr value and
select Save.
If max doses/hr is not visible on the home screen,
your CADD
not to require it. Instead you may be required to
program the hourly limit.
NOTE: If the hourly limit and the max doses/hr
settings are not displayed on your programming
screen, your facility's CADD
administrator has chosen not to require a delivery
limit beyond the PCA lockout.
Programming and Operation
®
-Solis system administrator has chosen
®
-Solis system administrator has chosen
®
-Solis system
Press here to Save
Press here to Save
31