Maximum Delivery Rate (combined bolus and
continuous)
40 to 175 mL/hr in increments of 1 mL.
Key Beeps
On or off.
Res Vol Low Trip Point
1 to 999 mL in increments of 1 mL.
Res Vol Empty Alarm
Insistent and one time only or non-insistent
and repeating.
Pump Stopped Alarm
Informational or high priority.
Air Detector
Turned on or turned off.
Air Detector Sensitivity
Low Sensitivity - Single bubble greater than
400 μL .
High Sensitivity - Single bubble greater than
150 μL .
Alarm Volume
High, medium, or low.
PM (Preventative Maintenance) Reminder
Interval: 1 to 24 months in 1 month increments.
Enable: On or off.
Custom Keypad Code
001 to 899 in increments of 1.
Custom Clinician Code
001 to 899 in increments of 1.
Custom Administrator Code
001 to 899 in increments of 1.
Date Format
US standard (mm/dd/yy) or European standard
(dd/mm/yy).
Time
00:00 to 23:59.
Downstream Occlusion Sensitivity
High Sensitivity - once the high pressure alarm
threshold is reached, the downstream occlusion
alarm is triggered immediately.
Low Sensitivity - once the high pressure alarm
threshold is reached, the downstream occlusion
alarm is delayed for 2 seconds, this allows for
References and Troubleshooting
the pressure to settle before a possible alarm
(if the pressure settles below the high pressure
alarm threshold before the 2 second delay is
complete, the alarm will not happen).
Upstream Occlusion Sensor
On or off.
NOTE: The upstream occlusion sensor
is automatically disabled during use with
medication cassette reservoirs.
Electromagnetic Emissions and Immunity
Declarations
Guidance and manufacturer's declaration - electromagnetic emissions
The CADD®-Solis pump is intended for use in the electromagnetic environment specified
below. The customer or the user of the CADD®-Solis pump should assure that it is used in such
an environment.
Emissions test
Compliance
RF emissions
Group 1
CISPR 11
RF emissions
Class B
CISPR 11
Harmonic emissions
Not applicable
IEC 61000-3-2
Voltage fluctuations/
Not applicable
flicker emissions
IEC 61000-3-3
Compliance using:
1. 100-240 VAC 50/60Hz to 7VDC switching
desktop AC adapter, with an AC power cord
length of 1.8 m (6 feet).
2. Rechargeable battery pack.
3. Remote dose cord with a length of 152 cm ± 5
cm (60 in. ± 2 in.).
4. USB cable length of less than 2 m (6.5 feet).
WARNING:
The use of power supplies and a remote
dose cord other than those listed in the
electromagnetic emissions declaration may
result in increased emissions or decreased
immunity of the pump.
The pump should not be used adjacent to or
stacked with other equipment. If adjacent
or stacked use is necessary, the user should
verify normal operation of the pump in the
configuration in which it is to be used.
Electromagnetic environment - guidance
The Pump uses RF energy only for its internal
function. Therefore, its RF emissions are very low
and are not likely to cause any interference in
nearby electronic equipment.
The Pump is suitable for use in all establishments,
including domestic establishments and those
directly connected to the public low-voltage power
supply network that supplies buildings used for
domestic purposes.