Pca Dose Graph; Delivery History And Pie Chart; Delivery Log; Event Log - Smiths Medical CADD-Solis Operator's Manual

Ambulatory infusion pump
Hide thumbs Also See for CADD-Solis:
Table of Contents

Advertisement

Programming and Operation

PCA Dose Graph

This screen displays the
number of doses attempted
and given in 30 minute
increments starting from
the current time to 8 hours
in the past or to the start of
a new protocol or patient.
In this example, the patient has attempted 7 doses
and 2 of those doses were given.
This is a good place to review the number of
attempted doses for a particular time frame.
The PCA dose graph can be viewed at any time,
with the pump running or stopped.
To view the PCA dose graph:
1. From the reports menu,
press U or D to
choose PCA Dose
Graph and press S.

Delivery History and Pie Chart

The delivery history and pie
chart is a pie chart view
of the total given over a
specified time frame or to
the start of a new patient or
protocol. The time frame
can be adjusted in various
intervals from 30 minutes
to 7 days.
Press U or D to view the time chart in the
various time frames. This will provide a quick
review of the methods of delivery over the interval
selected.
The delivery history is displayed in the units for the
current protocol.
The delivery history and pie chart can be viewed at
any time, with the pump running or stopped.
To view the delivery history and pie chart:
1. From the reports
menu, press U or D
to choose Delivery
History and Pie Chart
and press S.
28

Delivery Log

The delivery log is a subset of the event log and
contains information having to do specifically with
delivery events. Delivery log information includes:
PCA dose deliveries.
Clinician boluses.
Changes to the patient specific parameters
(including continuous rate, PCA dose, PCA
dose lockout, delivery limit).
Manually stopping a PCA dose and/or a
clinician bolus.
Starting a new protocol.
Pump started, stopped, powered up, and
powered down.
The delivery log is maintained by the pump, and
displays all entries since the last time a new patient
was started. The delivery log can be viewed at any
time, with the pump running or stopped.
To view the delivery log:
1. From the reports menu,
press U or D to
choose Delivery Log
and press S.
While viewing the delivery log, you may quickly
scroll from the oldest to newest by selecting Show
Oldest or Show Newest.

Event Log

The event log records the following types of events:
hourly given totals, dose delivery, alarms, error
codes, power source changes, cassette changes,
protocol library changes, and changes to pump
programming or settings. The pump records the
time and date of each event, and lists events in order,
with the most recent at the bottom of the screen
through the last 5000 events.
The event log can be viewed at any time, with the
pump running or stopped.

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents