2.4.5. Process Temperature Min.
The minimum recorded temperature of the internal electronics.
Open the menu View – Peak Values and select the tab Temperature. If necessary,
click on the reset button, and download to the device.
2.4.6. Process Temperature Max.
The maximum recorded temperature of the internal electronics.
Open the menu View – Peak Values and select the tab Temperature. If necessary,
click on the Reset button, and download to the device.

2.5. Linearization

Carries out a volume conversion from a level value.
2.5.1. PV (volume/level) Units
Select units for either volume or level.
2.5.2. Linearization Type/Vessel Shape
Defines the vessel shape and allows the LR 200 to calculate volume instead of
level. If "none" is selected, no volume conversion is performed. Select the vessel
shape matching the monitored vessel or reservoir.
*
20
21
Page 92
SITRANS LR 200 (PROFIBUS PA) – INSTRUCTION MANUAL
Vessel Shape
no volume calculation
----
required
flat end horizontal cylinder linearization volume
sphere
Description
N/A
linearization volume
Also required
7ML19985HP01