3 Block MEASURED VARIABLES
18
3.3
Group SPECIAL UNITS
3.3.1
Function group ARBITRARY UNIT
MEASURED VARIABLES
A
MEASURED VARIABLES
Use this function group to define an arbitrary unit for the flow rate variable.
TEXT ARBITRARY
Use this function to enter a text for the selectable volume (flow) unit. You define only the
VOLUME
text, the unit of time is provided from a choice of options (s, min, h, day).
(0602)
User input:
xxxxxxx (max. 4 characters)
Valid characters are A–Z, 0–9, +, –, decimal point, white space or underscore
Factory setting:
_ _ _ _ (No text)
Example:
If your text entry is "GLAS", this text string appears on the display complete with the unit
of time, e.g. "GLAS/min":
GLAS = Volume (text input)
GLAS / min = Volume flow as shown (on the display)
FACTOR ARBITRARY
Use this function to define a quantity factor (without time) for the free selectable unit.
VOLUME
The volume unit on which this factor is based is one liter.
(0603)
User input:
7-digit floating-point number
Factory setting:
1
Reference quantity:
Liter
Example:
The volume of a glass is 0.5 l
User input: 2
Device functions Proline Prosonic Flow 93
⇒
MEASURING VALUES
⇓
SYSTEM UNITS
⇓
SPECIAL UNITS
Functional description
➞
SPECIAL UNITS
→
2 glasses = 1 liter
AAA
ACA
⇒
AEA
ARBITRARY UNIT
➞
ARBITRARY UNIT
Endress+Hauser
060