3.2 Installing the Instrument
Installation Position
3-4
Desktop
Place the instrument on a flat, even surface as shown in the figure below. If the
instrument is installed in a horizontal position, rubber stoppers can be attached to the
feet to prevent the instrument from sliding. One set of rubber feet (two feet) are included
in the package. Do not install the instrument in positions other than the left two positions
indicated below.
Mounting in a Rack
To mount the instrument in a rack, use the separately sold rack mount kit.
Part Name
Rack mount kit (for single mount)
Rack mount kit (for multiple mount)
Rack mount kit (for single mount)
Rack mount kit (for multiple mount)
An outline of the attachment procedures is given below. For detailed procedures, see the
instructions that are included with the rack mount kit.
1.
Remove the handles from the left side of the instrument.
2.
Remove the four feet from the bottom of the instrument.
3.
Remove the two plastic rivets and the four seals covering the rack mount
attachment holes on each side of the instrument near the front.
4.
Place seals over the feet and handle attachment holes.
5.
Attach the rack mount kit.
6.
Mount the instrument on the rack.
M A
X
EX
T
M A
X
CL K
88)
LIN
VID EO-
AC
K/
OU T
T
(VG
A)
ET HE
RN ET
100
10 0
BA SE
LIN
-TX
K
ON
LY
US B
M IN
AL
A X
S
TO
CA
T
II
100
50
-24
0 V
VA
AC
MA
X 50
/60
Hz
Note
When mounting the instrument to a rack, allow at least 20 mm of space around the inlet and
vent holes to prevent internal heating.
Make sure to provide adequate support from the bottom of the instrument while avoiding
blocking the inlet and vent holes.
Model
Note
751533-E4
for EIA
751534-E4
for EIA
751533-J4
For JIS
751534-J4
For JIS
IM 760201-01E