YOKOGAWA WT500 User Manual page 47

Power analyzer
Hide thumbs Also See for WT500:
Table of Contents

Advertisement

2.6 Integration
2-24
Continuous Integration Mode
Integration continues for a specified amount of time, and then resets and starts again.
Integration repeats until you press STOP. If the integrated value reaches its maximum or
minimum displayable value before the specified amount of time has elapsed, integration
is stopped, and the integration time and integrated value at that point are held.
Timer
Timer
value
value
Start
Real-Time Normal Integration Mode
You specify the dates and times when integration will start and end. When the specified
stop time is reached, or when the integrated value reaches its maximum or minimum
displayable value, integration is stopped, and the integration time and integrated value at
that point are held.
Start
date/time
Hold
Hold
Timer
value
Press
STOP
Reset
Hold
Hold
End
Reset
date/time