Explanation
IM 760201-01E
For an explanation of the integration feature, see section 2.6.
For the WT500 to perform integration, you must set the parameters such as the
integration mode and the integration timer, and then start integration.
Selecting Normal or Continuous Mode
There are two integration modes that allow you to set a length of time (using the timer)
for which integration will take place. They are listed below.
• Normal Integration Mode
Integration continues until (1) the specified amount of time (which is determined by the
timer setting) elapses, (2) you press START/STOP, or (3) the maximum or minimum
displayable integrated value (see section 5.6 for details) is reached. When integration
stops, the integration time and integrated value are held.
• Continuous Integration Mode
Integration continues for a specified amount of time, and then resets and starts again.
Integration repeats until you press START/STOP. If the integrated value reaches its
maximum or minimum displayable value before the specified amount of time has
elapsed, integration is stopped, and the integration time and integrated value at that
point are held.
Setting the Integration Timer
• You can set the timer hour, minute, and second within the following range.
00000 : 00 : 00 to 10000 : 00 : 00
Note
If you set the integration timer to 00000 : 00 : 00 in normal mode, the WT500 will integrate in
manual mode (see sections 2.6 and 5.7 for details).
For information about integration, such as how to start, stop, and reset it, see the
explanations in section 5.6.