YOKOGAWA WT500 User Manual page 187

Power analyzer
Hide thumbs Also See for WT500:
Table of Contents

Advertisement

IM 760201-01E
6.7 Displaying Bar Graphs and Making Cursor Measurements
Measuring with Cursors
4.
Press SHIFT+FORM(CURSOR) to display the Cursor menu.
• Selecting Whether or Not to Use Cursor Measurement
5.
Use the cursor keys to select Bar Cursor.
6.
Press SET to select ON or OFF. If you select ON, the results of cursor
measurement are displayed.
• Moving the Cursors
7.
Use the cursor keys to select Position C1 + or Position C2 x.
8.
Press SET to display the cursor position selection box for Position C1 + or
Position C2 x.
9.
Use the cursor keys to set the position in terms of the harmonic order.
• Select Position C1 + to move the position of cursor +.
• Select Position C2 x to move the position of cursor x.
10.
Press SET or ESC to close the cursor position selection box.
• Linking Cursors
11.
Use the cursor keys to select Linkage.
12.
Press SET. The linkage indicator lights and the cursor settings are linked.
When Linkage is enabled, the distance between cursor + and cursor x is maintained. When
you change the position of one cursor, the position of the other cursor also changes.
Linkage indicator