YOKOGAWA WT500 User Manual page 150

Power analyzer
Hide thumbs Also See for WT500:
Table of Contents

Advertisement

5.8 Setting Normal or Continuous Integration
5-24
Setting the Integration Timer
8.
Use the cursor keys to select one of the Integ Timer boxes (hour, minute, or
second).
9.
Press SET to display an entry box.
10.
Use the cursor keys to set the hour, minute, or second that you selected in step 8.
11.
Press SET or ESC to close the entry box.
12.
Repeat steps 8 to 11 to set the hour, minute, and second.
13.
Press ESC twice to close the entry box and the setup menu.
Integrating in Normal or Continuous Integration Mode
• Starting Integration
14.
Press START/STOP. The integration indicator changes to Start (displayed in
green), and integration starts.
• Holding Integration
15.
Press HOLD. The HOLD key lights, and the numeric data display is held.
Integration continues.
• Un-Holding
16.
Press HOLD when the display is held. The HOLD key is no longer lighted, and
the numeric data display is updated. While the display is held, you can execute a
single measurement by pressing SHIFT+HOLD (SINGLE) to update the display.
• Stopping Integration
17.
Press START/STOP. The integration indicator changes to Stop (displayed in
red). The integration time and integrated value are held. If you press START/
STOP before the integration timer expires and then press START/STOP again,
integration will restart and continue until the timer expires.
In normal integration mode, once the integration timer expires, the integration indicator
switches from Start, displayed in green, to TimeUp, displayed in red. The integration time
and integrated value are held.
In continuous integration mode, once the integration timer expires, the integration time and
integrated value are automatically reset, and integration continues until you press START/
STOP.
• Resetting Integration
18.
Press SHIFT+START/STOP (RESET). The integration indicator changes to
Reset, the integration time and integrated value are reset, and their displays
change to "-----------" (no data).