Integration Mode (See sections 5.6 and 5.7 for operating instructions)
IM 760201-01E
The integration feature has five modes: manual integration mode, normal integration
mode, continuous integration mode, real-time integration mode, and real-time continuous
integration mode.
Manual Integration Mode
Integration continues from when you start it until you stop it. However, when the
integration time reaches its maximum (10000 hours) or the integrated value reaches
its maximum or minimum displayable value (see section 5.6 for details), integration is
stopped, and the integration time and integrated value at that point are held.
Start
When STOP is
pressed or the
maximum
integration
time is reached
Normal Integration Mode
Integration continues for a specified amount of time (which is determined by the timer
setting), and then stops. However, if the integrated value reaches its maximum or
minimum displayable value before the specified amount of time has elapsed, integration
is stopped, and the integration time and integrated value at that point are held.
Timer value
Start
Hold
Hold
Stop
Reset
Start
Hold
Hold
Reset
2.6 Integration
Hold
When the
integrated value
upper limit is reached
Hold
Reset