Run Option

Auto Export and Auto Record status:
Multi data display ON
Run Option button opens the Run Option dialog box used to change the status of the automatic data
record function, the automatic data export function, and the multiple data display function.
If the Record Test Result Data Automatically check box is checked, both or one of the automatic data
record function and the automatic data export function will be set to ON. To specify the function
status, see the next page.
If the Record Test Result Data Automatically check box is not checked, both functions are set to
OFF. In this status, both data record and data export are not performed. So the test execution time
will be reduced. Also, EasyEXPERT memorizes the last test result data. And it can be saved as a
test result record by clicking the Save Data button right next to the Run Option button. This status
will be useful for probe contact check, defect analysis, and so on.
The multiple data display function ON or OFF can be controlled by clicking the Multi Display
indicator or by using the Allocate Data Display for each test check box on the Run Option dialog
box.
About the multiple data display function:
If this function is OFF, the all test result data will be displayed on the singular Data Display window.
If this function is ON, the test result data of the same test setup name will be displayed on the
exclusive Data Display window and the test result data of the different test setup name will be
displayed on the new Data Display window.
or
or
Auto Export OFF and Auto Record OFF
Multi data display OFF
12-4
Miscellaneous Operations
Save Data button
appears if the last
test result data is in
the memory when
both Auto Export and
Auto Record are OFF,
and allows you to save
the last data as the
test result record.