Agilent Technologies B1500A Training Manual page 42

Semiconductor device analyzer
Hide thumbs Also See for B1500A:
Table of Contents

Advertisement

To Define Test Output
Display Parameters
This example sets:
X-Y Graph: ID-VG plot for Y1-X graph and GM-VG plot for Y2-X graph
List Display: VG, ID, and GM
Parameters: GM_MAX and VTH
All parameters must be defined in the Analysis Parameter Definition window shown in the previous
slide.
The X-Y Graph area is used to set the X axis and Y axis of the X-Y Graph Plot area on the Data
Display window.
•Name: Name of variable to plot on the X axis
•Scale: Linear or Log
•Min: Minimum value of the specified axis
•Max: Maximum value of the specified axis
The List Display area selects the variables to be listed on the List Display area of the Data Display
window. Up to 20 variables can be set.
The Parameters area selects the variables to be listed on the Parameters area of the Data Display
window. Up to 20 parameters can be set.
10-22
Module 10
Creating Your Test Definitions

Hide quick links: