Flash Memory – Endurance Test
NOR Erase Connection
Selector or ASU ch2
SMU3
PG3
This is an example setup to perform the erase operation of the NOR type flash memory cell. To
perform the erase operation, all of the selector/ASU channels must make the path to the PG and the
PG3 internal pulse switch must be opened. The pulse switch is also a solid state relay.
Note that the solid state relay in the selector is not used. The relay must be always close.
Selector or ASU ch3
Drain
Mech
Substrate
Mech
Source
Gate
13-27
SMU4
Mech
Solid
Mech
PG4
State
SMU1
Selector or ASU ch1
SMU2
Mech
Solid
Mech
PG2
State
Module 13