Programmed Operation; Creating A New Program - Beckman Coulter Avanti J-25 Instruction Manual

High-performance centrifuge
Table of Contents

Advertisement

Operation

PROGRAMMED OPERATION

CREATING A NEW PROGRAM

Action
1. Press [
].
OPT
2. Press [
] again to move the
OPT
cursor to the Setup: field.
Use the keypad to enter a
program number in this field.
3-14
NOTE
If a diagnostic condition causes the run to stop, a
diagnostic message appears and the red [
indicator light flashes until the rotor comes to a
complete stop. See Table 4-1 for diagnostic code
explanations and troubleshooting procedures.
The centrifuge internal memory can store up to thirty programs, each
containing up to two steps (each step is a set of run parameters).
Programs are retained in memory when the power is turned off.
Result
Default system parameters appear, the Prog(ram) Recall: field
flashes, and the message line says "Enter program number (1–30)."
A two-step program is created and is assigned the number you
entered. If you enter a number that is already assigned to a program,
you can either overwrite the program or select another program
number.
A message prompts you to enter parameters for step 1.
Rotor : JA-25.50
Speed : 0
Time :
00:00
Temp : 25°C
Accel : MAX
Prog Recall:
Enter parameters for the selected step
RPM
=
x G
Max Temp : 29°C
Shutdown? No
Decel : MAX
Setup: 1
Step #1
]
STOP
STEP1
STEP2
SAVE
EXIT

Hide quick links: