ACCEL
(Acceleration)
[
]
DECEL
(Deceleration)

PROGRAMMED OPERATION

[
]
PROG
(Program)
and
[
]
SAVE
The instrument has two acceleration profiles: maximum (MAX) and
slow. Press the [
] key repeatedly to toggle between the settings.
ACCEL
The run status indicator for the selected setting will light.
The instrument has three deceleration profiles: maximum, slow, and
no brake. Press the [
DECEL
settings. The run status indicator for the selected setting will light.
Up to nine sets of user-specified run conditions, called programs, may
be stored. Each program is stored by number and can be recalled by
using the keypad. Programs are retained in memory even when the
power is turned off.
You can select an existing program by pressing the program number
(1 to 9) on the keypad, or by pressing the [
the required set of parameters appears.
To create or modify a program:
1. Press [
]. If no program has been previously selected, a dash
PROG
will appear above the [
other displays.
2. Use the keypad to select a program number, which replaces the
dash in the display. Or, continue pressing [
program number appears. The program number blinks to indicate
that you are editing the program.
3. Use the parameter keys and keypad to enter the run settings. After
all parameters have been entered, press [
SAVE indicator will begin to blink to indicate that you may save
the program.
4. Press [
] and the program will be stored in memory under the
SAVE
selected number. Actual values will reappear in the displays.
] key repeatedly to toggle between the
] key repeatedly until
PROG
] key. Actual values remain in the
PROG
PROG
ENTER/RECALL
Run Preparation
] until the required
]. The
2-7