Beckman Coulter Avanti J-25 Instruction Manual page 19

High-performance centrifuge
Table of Contents

Advertisement

Specifications and Preinstallation Requirements
Rotor Profile and Description
JS-24.15 Swinging Bucket (6 place)
JS-13.1 Swinging Bucket (6 place)
JS-7.5 Swinging Bucket (4 place)
JS-5.9 Swinging Bucket (2 place)
JCF-Z Continuous Flow/Zonal Rotor
JE-6B Elutriation Rotor
a
Maximum speeds are based on a solution density of 1.2 g/mL with the following exceptions: the JA-18.1 rotor is rated for a density of 1.4 g/mL; the
JCF-Z rotor is rated for a density of 1.45 g/mL; and the JE-6B rotor is rated for a density of 3 g/mL.
b
Maximum speed for the JA-30.50 Ti rotor is 25 000 rpm in the Avanti J-25I centrifuge, and 30 000 rpm in the Avanti J-30I centrifuge.
c
Maximum speed for the JS-24.38 and JS-24.15 rotors is 10 000 rpm in the Avanti J-25I centrifuge, and 24 000 rpm in the Avanti J-30I centrifuge.
1-6
Max
RPM
10 000
r
= 171.3 mm
max
13 000
r
= 140 mm
max
r
= 165 mm
7 500
max
r
= 168.5 mm
5 900
max
20 000
r
6 000
max
standard
chamber = 125 mm
Sanderson
chamber = 126 mm
Max
RCF
Max
a
(× g)
Capacity
6 × 15 mL
c
19 200
6 × 50 mL
26 500
4 × 50 mL
10 400
6 570
10 microplates
4 deep-well plates
2 square-well plates
660 mL
(standard core)
1250 mL
(large core)
240 mL
(small core)
5 040
4.2 mL
(standard chamber)
5 080
4.8 mL
(Sanderson chamber)
Rotor
Manual
Number
J-TB-058
J-TB-036
J-TB-007
J-TB-092
JCFZ-IM
JE6B-IM

Hide quick links: