Entering Run Temperature - Beckman Coulter Avanti J-25 Instruction Manual

High-performance centrifuge
Table of Contents

Advertisement

Operation
Action
3. Press the next function key or
press [
ENTER

ENTERING RUN TEMPERATURE

Action
1. Press [
].
TEMP
2. Use the keypad to enter the
required temperature.
3. Press the [
TEMP
to move the cursor to the
Max Temp: field.
4. Use the keypad to enter a
different maximum tempera-
ture, or accept the default.
3-10
Result
After you start the run, the HOLD indicator illuminates in the TIME
], then [
].
display. The elapsed run time is displayed. The run continues until
START
[
STOP
Run temperature can be set from –20 to +40°C. If no value is entered,
the last-entered temperature is used.
Result
The Temp: field flashes.
Entered temperature appears in the Temp: field.
] key again
Entered temperature plus 4°C appears in the Max Temp: field, and
YES and NO soft keys appear. The default maximum temperature is
the set temperature plus 4°C. You can specify a different maximum
temperature, from +2°C above the set temperature to 44°C. You can
also specify whether or not you want the system to shut down if the
maximum temperature is reached. The centrifuge will always shut
down if the system temperature reaches 45°C.
] is pressed.
Rotor : JA-25.50
Speed : 25 000
RPM =
Time :
00:00
Temp:
- -
Accel : MAX
Decel : MAX
x G
Max Temp : - - °C