Check all J2 series rotors for drive pins, and do not use rotors without
drive pins in the Avanti J. To check for drive pins, hold the rotor up or
turn it on its side and look into the drive hole. If you do not see metal
pins in the hole, do not use the rotor in the Avanti J. Call your local
Beckman Coulter office for information on returning the rotor to the
factory for upgrading.
In Avanti J series centrifuges, the JA-18 rotor must be run with the
lid attached. See the JA-18 rotor manual (document J-TB-035) for
complete rotor usage instructions.
A special bracket and mounting hardware are required when the
JCF-Z Continuous Flow/Zonal Rotor is used. These parts are
contained in the Zonal Bracket Kit (part number 363843). Kit part
number 363843 includes a bracket, mounting hardware, and assembly
instructions. Kit 363843 is for use with Cole-Parmer tubing size 16
(6.4 mm [
1
/
-in.] O.D.), the tubing size most commonly used with the
4
JCF-Z rotor.
Two additional kits are available, one which contains adapters for
size 14 tubing (4.8 mm [
adapters for size 15 tubing (9.6 mm [
either of these sizes of tubing, you will need to purchase one of these
kits in addition to the basic kit (363843).
See
REPLACEMENT PARTS AND SUPPLIES
complete list of Zonal Bracket Kit part numbers and kit contents.