Error Condition; Firmware Revision - Picotest M3510A User Manual

6 1/2
Table of Contents

Advertisement

memory
How to set up the language
Do the procedure SHIFT + CONFIG > SYSTEM > ENTER > LANGUAGE >
ENTER > DEFAULT/COMPATIBLE > ENTER to select the language.

4.4.7 Error Condition

The error annunciator on the front panel display tells about the error
condition of the multimeter. If there are one or more syntax or hardware
errors found, the error annunciator will be lit to inform you. The
multimeter stores errors using the first-in-first-out (FIFO) order and it
records up to 20 errors in the error queue. The first error detected will be
the first error retrieved.
are listed in the following. For more ERROR information,
6 for the list of error messages.
The error annunciator will be turned off when you have read the
errors in the memory.
If there is no error found when you check the error queue, the
multimeter shows "NO ERRORS".
If there are more than 20 errors found, the last error in the error
queue will be replaced with "-350", means too many errors.
The error queue is cleared when the multimeter is turned off, or when
a *CLS command is received.
How to check the error queue
Do the procedure SHIFT + CONFIG > SYSTEM > ENTER > SCPI ERR >
ENTER to check the error message, such as "ERROR CODE" or "NO
ERRORS".

4.4.8 Firmware Revision

M3510A/11A has three microprocessors for various internal systems.
You can query the multimeter to determine which revision of firmware is
In addition, the general conditions of the ERROR