Initial Mode; Language - Picotest M3510A User Manual

6 1/2
Table of Contents

Advertisement

Default
The default band is 0.1%. The defined selection is stored in a volatile
memory and it will be cleared after the meter has been turned off.
How to enable/disable HOLD feature
1.
Press SHIFT + TRIGGER buttons. The HOLD anunciator will light up
when HOLD feature is enabled.
2.
To disable HOLD feature, press SHIFT + TRIGGER again.
How to adjust the sensitivity band
You can adjust the sensitivity band through either the front panel
operation or the remote interface operation.
Front Panel Operation
Do the procedure SHIFT + CONFIG > TRIG SYS > READ HOLD >
0.01%/0.1%/1.0%/10% to choose a desired band. Press ENTER on the
desired sensitivity band.

4.4.5 Initial Mode

The section contains two selections: "DEFAULT SET" and "SAVE DATA".
You can select "SAVE DATA" to save the current settings which will exist
after restarting next time. Or select "DEFAULT SET" to recall the factory
value after restarting M3510A/11A. The SAVE DATA is available on using
MEAS FUNC, RANGE, RESOLUTION, MATH FUNC & 2ND STATE.
How to save the settings
Do the procedure SHIFT + CONFIG > SYSTEM > ENTER > INIT MODE >
ENTER > SAVE DATA/DEFAULT SET > ENTER to save your current setting
or to resume the default setting.

4.4.6 Language

M3510A/11A supports two languages: DEFAULT (M3510A/11A) and
COMPATIBLE. When finished, the setting is stored in a non-volatile
80 80 80 80