a volatile memory. The default values for both upper and lower limits are
"0". This function is available to all except continuity and diode
measurements.
How to set the limits
You can set the limits or make a limit testing either through the front
panel or the remote interface operation.
Front Panel Operation
1.
Do the procedure SHIFT + NULL > LIMITS > ENTER > HIGH
LIMIT/LOW LIMIT > ENTER to access the setting.
2.
Use
and
◁
selection of LOW LIMIT or HIGH LIMIT.
3.
Use
and
◁
▷
or decrease the numbers to a desired value, then confirm the value
with ENTER button. Press ENTER on "RUN" to start measurement.
Note: To disable the limits function, you always need to go to Math
※
submenu and select "OFF" or press other measurement buttons. The
"MATH" annunciator should not be lit when disabled.
Remote Interface Operation
Use the following commands to enable the limits function or to set the
limits:
CALCulate:FUNCtion LIMit
CALCulate:STATe {OFF|ON}
CALCulate:STATe?
CALCulate:LIMit:LOWer {<value>|MINimum|MAXimum}
CALCulate:LIMit:UPPer {<value>|MINimum|MAXimum}
4.3.5 MX+B
Definition
This mathematical function multiplies a measurement reading (X) by a
specified scale factor (M) and add an offset (B) automatically. The
to toggle between options. Press ENTER on your
▷
to move through the digits and
△
71 71 71 71
and
to increase
▽