can't work at the 10A range.
Front Panel Operation
1. Choose a measurement function.
2. Press SHIFT + ESC buttons to select auto-range feature. Or use
and
buttons to manually select a range (MAN anunciator will lit
▽
when manually selecting a range).
Remote Interface Operation
Use the following commands from your PC terminal to set the range:
CONFigure:<function> <range>,<resolution>
MEASure:<function>? <range>,<resolution>
SENSe:<function>:RANGe <range>
SENSe:<function>:RANGe:AUTO {OFF|ON}
4.1.6 Rate (Integration Time)
Definition
Integration time is the period of time the analog-to-digital (A/D)
converter takes to sample the input signal. The integration time feature
allows you to optimize either the speed or the resolution, as well as the
noise rejection and the accuracy of the measurement. The unit of the
integration time is in PLC (power line cycles). One PLC for 60 Hz is 16.67
ms, and for 50 Hz is 20 ms.
There are 10 different integration times in M3510A/11A for you to select
from:
0.001,
0.006, 0.02, 0.06, 0.2, 0.6, 1, 2, 10 and 100 PLCs.
Note: Some following functions or instructions which the M3511A
※
doesn't support or provide are colored green.
Default
For DCV, DCI, and resistance measurement, the default integration time
is 1
PLC. The manual selected value is stored in a volatile memory and
0
the default value will be restored when the meter is power-off.