Range (Manual & Auto) - Picotest M3510A User Manual

6 1/2
Table of Contents

Advertisement

Default
The factory default for continuity threshold resistance is 10Ω. Your
selection is stored in a volatile memory and the default value will be
restored after the meter has been turned off.
How to set the threshold resistance
You can change the threshold resistance only through the front
panel.
1. Press CONT button and then CONFIG button.
2. Press ENTER button to access the threshold resistance settings.
3. Use
and
buttons to move through the digits and lower or
increase the number to your desired value by using
and
buttons.
4. Press ENTER button again to confirm the value.
4.1.5 Range (Manual & Auto)
Definition
When making measurements except CONT, DIODE, TCOUPL and
Temperature, M3510A/11A will automatically choose a range for you, or
you can select an appropriate range manually. The difference between
auto-range and manual-range is the settling time. Auto-range is a
convenient way for you, but manual-range can usually speed up the
process.
If the input signal is beyond allowed range, an "OVLD" message will be
shown on the display. The thresholds of maximum/minimum readings
for each range are 120 % of the range for maximum and 10% of the
range for minimum.
Default
The default is auto-range. The manual selected range is stored in volatile
memory and the default will be restored when the meter is power-off.
Please refer to Table 2-1 on page 23 for the factory default range.
How to set the auto/manual range:
You can set the auto/manual range either through the front panel
operation or through the remote interface operation. The autorange
51 51 51 51