Trigger Setting - Picotest M3510A User Manual

6 1/2
Table of Contents

Advertisement

Output
Approximately
2 µ
s s s s
Figure 4-4
Remote Interface Operation (By software or internal trigger)
By software trigger: The software trigger is similar to the single
triggering, but instead of using the TRIGGER button on the front panel,
you send a command from your PC to the multimeter to generate an
event. For using software trigger, use the following command from your
PC terminal to set trigger source:
TRIGger:SOURce BUS
By internal trigger: The internal trigger is the default trigger mode for
the Remote Interface Operation. In the internal trigger mode, a trigger
signal will be issued whenever the multimeter is in the wait-for-trigger
state. To set the internal trigger, use the following command from your
PC terminal:
TRIGger:SOURce IMMediate

4.2.3 Trigger Setting

In M3510A/11A, you can specify a variety of trigger settings including
the number of samples per trigger, the number of triggers per event,
reading hold, and the trigger delay for your measurements.
A. Number of samples on each trigger
By default, M3510A/11A takes only one reading on each trigger, but you
can instruct the multimeter to take specific number (up to 2000) of
readings each time it receives a trigger. The defined input setting is
stored in a volatile memory which will be cleared after the meter has
63 63 63 63