frequency measurements return "0".
CONFigure:PERiod {<range>|MIN|MAX|DEF},{<resolution>|MIN|MAX|DEF}
Preset and configure a period measurement with the specified range and
resolution. This command does not initiate the measurement. For period
measurements, the meter uses only one "range" for all inputs between 0.33
seconds and 3.3 sec. With no input signal applied, period measurements return "0".
CONFigure:CAPacitance {<range>|MIN|MAX|DEF},{<resolution>|MIN|MAX|DEF}
Preset and configure the multimeter for capacitance measurements with the
specified range and resolution. This command does not initiate the
measurement.
CONFigure:CONTinuity
Preset and configure for a continuity measurement. This command does not
initiate the measurement. The range and resolution are fixed at 1kΩ and 5½
digits respectively.
CONFigure:DIODe
Preset and configure for a diode measurement. This command does not initiate
the measurement. The range and resolution are fixed at 1Vdc with 1mA current
source and 5½ digits respectively.
CONFigure:TEMPerature
Preset and configure for a RTD measurement. This command does not initiate
the measurement. The resolution is fixed at 6½ digits.
CONFigure:TCOuple
Preset and configure for a thermocouple measurement. This command does not
initiate the measurement. The resolution is fixed at 6½ digits.
CONFigure?
Query a recent condition of function, and return a quoted string.
C.5 The Measurement Configuration Command
[SENSe:]FUNCtion "<function>"